2.2.6 Power Equipment
2.2.6.1 Radio‐Frequency Instrumentation
Monopolar or bipolar radio‐frequency instrumentation is used to cut tissue and cauterize bleeding vessels. Standard radio‐frequency systems for traditional open surgery are usable for minimally invasive surgery with appropriate monopolar or bipolar cords connected to minimally invasive surgery instruments. Minimally invasive surgery hand instruments have a connector pin for monopolar electrosurgery (Figure 2.12). Adaptation of monopolar electrosurgery units for bipolar function is also possible with bipolar MIS instruments connected to the radio‐frequency generator with a special power cord that allows placement of one connector to the active socket and the other connector to the ground socket (Figure 2.13). This drives the current between the two sides of the instrument creating a bipolar instrument. Bipolar generators are also used with a power cord for this application. Monopolar and bipolar radiofrequency is capable of sealing vessels up to 3 mm in diameter and is sufficient for most minimally invasive surgery in small animal practice.
Figure 2.12 A minimally invasive surgery instrument with a connection post for monopolar electrosurgery. The connection post is the silver split projection at the top right of the pistol grip handle. This post is attached to the active side of the electrosurgery generator using a standard open‐surgery grounding plate for the inactive or ground connection.
(Source: Photo courtesy of Karl Storz: ©Karl Storz SE & Co KG, Germany.)
Figure 2.13 A minimally invasive surgery bipolar vessel sealing and cutting instrument, Karl Storz 5 mm diameter RobiPlus, with a connection post for bipolar cautery seen as a black cylinder projecting from the upper right of the instrument handle. This instrument is used with a monopolar generator and a power cord with two plugs, one that goes into the active socket, and the other goes into the ground socket. This instrument also works with a bipolar generator using a proper cord for this application.
(Source: Photo courtesy of Karl Storz: ©Karl Storz SE & Co KG, Germany.)
2.2.6.2 Vessel Sealing Devices
Systems capable of sealing larger blood vessels, up to 7 mm diameter, were developed for minimally invasive surgery use in human patients. These devices use tissue fusion technology that applies low voltage with high current across a bipolar instrument while measuring tissue impedance (resistance) to achieve a tissue weld sealing the blood or lymphatic vessel. The original vessel sealing device using this technology was the LigaSure and has been the workhorse in small animal practice for over 20 years. Originally introduced for minimally invasive surgery with sealing and cutting capability in one instrument, it also became a standard instrument for open surgery with handpieces developed specifically for open surgery. A second‐generation device produced by the same company and the standard today is the ForceTriad (Figure 2.14) with advanced technology for faster function but little change in sealed vessel properties. The ForceTriad also has monopolar and bipolar modules. A wide variety of instruments are available for minimally invasive surgery in 5 and 10 mm diameters (Figure 2.15) and in lengths from 20 to 37 cm with sealing and cutting capabilities. There is also selection of open surgery instruments just for sealing (Figure 2.16) and sealing plus cutting in one instrument (Figure 2.17). A third‐generation device is now available, but this step has incorporated circuitry to prevent reuse of the handpieces currently making it unsuitable for small animal practice. The company has indicated that they are working on a modification to allow reuse of the handpieces in veterinary medicine. Competing companies have developed similar vessel sealing devices that are now available and are worth considering for use in small animal practice.
Figure 2.14 The ForceTriad vessel sealing device with standard monopolar radiofrequency on the left, monopolar or bipolar radiofrequency in the center, and vessel sealing on the right side of the control box. All three modalities can be connected and active simultaneously.
(Source: Photo courtesy of Covidien Division of Medtronic's.)
Figure 2.15 A 10 mm diameter 37 cm long “Atlas” vessel sealing instrument for minimally invasive surgery with a round shaft to provide a seal for maintaining insufflation during laparoscopy. This is the workhorse for minimally invasive surgery in small animal practice. Multiple‐length 10 mm instruments and similar 5 mm diameter instruments are available for use with this system.
(Source: Photo courtesy of Covidien Division of Medtronic's.)
Figure 2.16 An open‐surgery handpiece for use with the ForceTriad that seals but does not cut.
(Source: Photo courtesy