Название | Point-of-Care Ultrasound Techniques for the Small Animal Practitioner |
---|---|
Автор произведения | Группа авторов |
Жанр | Биология |
Серия | |
Издательство | Биология |
Год выпуска | 0 |
isbn | 9781119461029 |
Default back to original settings. If you still cannot figure out why the image looks so wrong, push the B‐mode button first and then go to presets and change to a different preset or return to the original preset, as many machines will default back to the original settings. The reality is sometimes you have no idea what you did to make everything look so bad.
See Chapter 4 for additional explanation and examples.
Not Knowing How to Freeze and Roll the Cine Ball
You will need to know how to freeze the image on the screen. This is especially helpful when you have short‐duration images of the region of interest. Examples include air (lung) interference of cardiac views in a panting patient, the tachycardic patient with normal higher heart rates (cats) in which the eyeball method is error prone; or you caught a quick glimpse of a structure you want to measure. Freezing the image on the screen and then rolling the cine ball through preceding frames allows you to evaluate better than in real time.
Figure 5.17. Slice‐thickness artifact at the AFAST CC view. (A) and (B) are the same images within the same cine loop that have been extracted to show how much gain settings affect the image. In (A) is a slice‐thickness artifact that is circled in (C). In (B) the gain is turned down, eliminating the artifact (true sediment would persist in gravity‐dependent regions). Color flow Doppler is another technique used to look for flow (mass versus thrombus versus sediment versus artifact). (C) and (D) are labeled images of (A) and (B) respectively. Note how the loop of small intestine outside the lumen of the urinary bladder could mimic a bladder stone (cystic calculi) with a hyperechoic line (reflective air) and clean shadowing through the far‐field. Note also that the screen provides a lot of information including a depth of 10 cm, an abdominal preset (Abv) in the upper right. Note, there is uniquely no focus cursor on machines from this manufacturer (SonoSite). Arrows indicate the body wall in the CC pouch. SI, small intestine; UB, urinary bladder.
Source: Courtesy of Dr Gregory Lisciandro, Hill Country Veterinary Specialists and FASTVet.com, Spicewood, TX.
Image Interpretation
Lacking Awareness of Most Common Locations for Artifacts
In Chapter 3, Table 3.1 lists the common ultrasound artifacts and the most common locations and examples for each. Briefly, we will list here some strategies to create better awareness of these artifacts, especially during the Global FAST approach. This leads to another major point…
Saving images is important not only for archiving patient information but for learning by reviewing studies and being able to compare to other modalities, future studies, and clinical course and outcome. For learning artifacts and how your machine interprets them, you can go through studies playing the “artifact game.” For example, look at the AFAST views for mirror image artifact at the DH view on the pleural cavity side of the diaphragm, the CC view and the other side of the abdominal wall, and TFAST echo views for mirror image artifact on the far side of the pericardium. This can also be done with other artifacts in their most common locations, such as edge shadowing off the curved surface of the stomach wall, renal cortex, urinary bladder, etc.
Mirror Image
This most commonly occurs at the diaphragm but the most problematic is the mirroring of the gallbladder into the thorax, especially when it’s a partial mirror image. In this case, the gallbladder can mimic pleural and pericardial effusion to the hasty sonographer.
Other places for mirror image, where generally a strong air–soft tissue interface exists, are on the far side of the heart when aerated lung is in direct opposition to the pericardium, and the urinary bladder against the colon or thin body wall. See Chapter 3 for more explanation and examples.
Side‐lobe, Slice‐thickness
This artifact occurs where there is a curve within a fluid‐filled structure and is especially associated with the urinary bladder and gallbladder mimicking sediment and a mass. To differentiate artifact from sediment, consider where the gravity‐dependent region is within the lumen of these structures. If “sediment” appears in nongravity‐dependent regions, you should consider the possibility of artifact. Other techniques to differentiate artifact from true pathology include eliminating the artifact by turning down the gain and ballottement or changing the patient’s position. Finally, some hasty sonographers will misinterpret an artifact for a mass. Simply place color flow Doppler on the suspect region for differentiation because a mass generally has pulsatile blood flow and an artifact (and thrombus) does not. See Chapter 3 for more explanation and examples.
Edge Shadowing
The edge shadowing artifact is created when the ultrasound beam strikes a curved surface, creating a dark shadow that extends from that curved surface through the far‐field. Edge shadowing most commonly occurs at the curved surfaces of stomach wall, renal cortex, gallbladder and urinary bladder and can be mistaken for free fluid along the stomach wall and renal cortex or defects in the wall of the gallbladder and urinary bladder, the latter especially when ascites is present on the AFAST CC view. See Chapter 3 for more explanation and examples.
Pearls, Pitfalls, and The Final Say
The paradigm change of POCUS and FAST rests on expediting the learning process and we hope that this chapter will help accelerate your learning process by highlighting factors in a different way from other chapters in this textbook. The material in the chapter is based on training more than 1000 veterinarians in these techniques over the past 15 years as well as leading in clinical studies.
Take the time to go through the major knobs on the machine ‐ depth, gain (TGC and overall), focus cursor, frequency, and preset. Know where these buttons are located and play with images to get used to how to efficiently adjust their settings. Using stickers on your machine is another way to train yourself on where these buttons are located.
Use our algorithm for trouble shooting. By doing so, you will have a way to efficiently review the most common image optimization features of your machine. Depth, frequency, preset, focus cursor, gain (TGC and overall), probe–skin contact, direction of the beam, then default back to original settings.
Save images and then go through the views and play the artifact game; look through AFAST studies and see how many mirror image artifacts you see and then choose another artifact and do the same.
References
1 Boysen SR, Rozanski EA, Tidwell AS, et al. 2004. Evaluation of a focused assessment with sonography for trauma protocol to detect free abdominal fluid in dogs involved in motor vehicle accidents. J Am Vet Med Assoc 225(8):1198–1204.
2 Walker C, Mohabir PK. 2014. Orientation. In: Point‐of Care Ultrasound, edited by Soni N, Arntfield R, Koy P. Philadelphia: