Magnetic Resonance Microscopy. Группа авторов

Читать онлайн.
Название Magnetic Resonance Microscopy
Автор произведения Группа авторов
Жанр Химия
Серия
Издательство Химия
Год выпуска 0
isbn 9783527827251



Скачать книгу

for transverse magnetic (TM) and hybrid (HEM) mode types.

      Figure 2.2 Resonant mode field estimation: part-by-part decomposition of the computation volume, case of the disk resonator.

      For example, the field expression remains unchanged within the dielectric disk, while outside it is written as in Equation 2.6, with f a decaying function depending on the computation volume boundaries, Kn the modified Bessel function of the first kind and order n, and ν the radial wavenumber in the region where the EM field is radially decaying:

      2.3.2 Power Loss Contributions in a Ceramic Probe

      While operating in different regimes, the loss phenomenon is the same in the ceramic material and in the sample: it is energy dissipated as heat within complex permittivity materials immersed in an electromagnetic field [19]. In practice, these power losses are expressed as the integral over the object volume V of the power loss density, which involves two local variables: the imaginary part of the material permittivity and the electric field intensity. The power losses in a material of complex permittivity real quantities) are expressed in Equation 2.7.

      The imaginary part of the permittivity is equal to the real part of the permittivity, ϵ0 the vacuum permittivity, ϵr the material relative permittivity, and tan δ the total loss tangent. Materials are usually described by the quantities that can be measured in experiments, the real part of the permittivity, and the total loss tangent. The latter quantifies the effects of two distinct phenomena responsible for EM power losses in the material: bound charge polarization and electric conduction when the material has a microscopic nonzero free electron density [20]. As biological samples contain a significant proportion of water, the conductivity of the sample is dominant because of dissociated ions available in the water solution, and the losses depend on the electrical conductivity σ:

P subscript text loss end text end subscript superscript text sample end text end superscript equals 1 half triple integral subscript V sigma subscript text sample end text end subscript vertical line E vertical line squared d v

      The ceramic probe is modeled as a ceramic ring resonator (inner radius rh, outer radius rd, height L, relative permittivity ϵr, loss tangent tan δ) filled with a cylindrical biological sample (radius rh, height L, and electrical conductivity σsample). With the theoretical insight about the TE01δ mode field distribution provided in Section 2.3.1, it is possible, as detailed in [21], to develop an analytical expression for the power losses in the ceramic probe at the cost of some approximations:

       The field distribution used to express the dielectric resonator losses is that of a lossless resonator because losses in the ceramic are considered small (tan δ ≪ 10−1).

       The field distribution of the ring resonator is assumed equal to that of the corresponding disk without field leakages at the lateral boundaries.

      With these assumptions, the power losses expression reduces to Equation 2.8 with the axial wavenumber ky known from the mode study, the maximum magnetic field amplitude in the disk resonator, and τ a penalty coefficient accounting for the field decreasing in the sample compared with the disk field distribution, estimated from the knowledge of both the ring and disk field distributions.