Название | Gastroenterological Endoscopy |
---|---|
Автор произведения | Группа авторов |
Жанр | Медицина |
Серия | |
Издательство | Медицина |
Год выпуска | 0 |
isbn | 9783131470133 |
One variable that is entirely under the control of the operator is duration (time) of current application, which can impact the final outcome to a large degree. Taking into consideration that the total amount of energy (J) delivered equals power (W) ×time (s), how fast or slow the power is being deposited becomes important. The outcome of delivering 50 W of power for 2 seconds is quite different than that of delivering 20 W for 5 seconds, even though the total amount of heat energy delivered (100 J) is the same. Furthermore, if the time of application and power settings remain the same, either a continuous or modulated waveform will deliver an identical amount of total energy but with a very different outcome (see the section Electrosurgical Units and Waveforms).
Fig. 7.3 Broad power-to-resistance curves suitable for snare polypectomy.
Fig. 7.4 Narrow power-to-resistance curves suitable for bipolar hemostasis.
7.2.2 Monopolar versus Bipolar Circuit
The terms monopolar and bipolar refer to the manner in which the electrosurgical circuit is completed. In a monopolar circuit, RF current oscillates from the active electrode (e.g., polypectomy snare) through the patient’s body via the path of least resistance to the inactive or dispersive electrode (electrosurgical pad), and returns to the ESU to complete the circuit (
In a bipolar circuit, the active and return electrodes are in close proximity to each other, as illustrated by the bipolar hemostatic probe (e.g., Gold Probe, Boston Scientific Inc., Natick, MA) (
Fig. 7.5 Schematic of monopolar circuit. Current flows from the active electrode (e.g., snare) through the patient’s body to the dispersive electrode (pad) placed on the patient’s skin.
Fig. 7.6 Schematic of bipolar circuit. Bipolar hemostasis probe with active and return electrodes closely spaced at the probe’s tip (inset).
7.3 Electrosurgical Units and Waveforms
ESUs enable the user to select from a menu of current waveforms a particular output, such as “Cut,” “Coag,” and “Blend.” An understanding of the waveform selected for a given ESU is important since manufacturer’s labeling of the outputs is not standardized and can be misleading (
ESUs produce current outputs ranging from low-voltage, continuous sinusoidal waves to interrupted (modulated) waveforms with much higher voltages (
Fig. 7.7 Schematic of current waveforms (see text for details). (Repoduced with permission from Macmillan publishers Ltd: Am J Gastroenterol, copyright 2009.3)
Modulated or interrupted waveforms heat tissue more slowly than continuous waves and are designed for outputs intended to produce varying amounts of coagulation, ablation, and hemostasis. By interrupting the current flow, tissue has a chance to cool, and the proportion of cells that desiccate without bursting increases. These modulated waveforms are recognized by particular labels, such as “Blend,” “Forced Coag,” and “Coag.” Peak voltages for modulated waveforms range from well over 200 Vp to 4,500 Vp, and even higher for noncontact fulguration modes. High-voltage modulated waveforms are used to ionize argon gas during noncontact argon plasma coagulation (APC).
One should not assume that a specific label, such as “Blend,” from one ESU corresponds to the same waveform output from another ESU with the same name. When reporting ESU settings in research studies and publications, quantitative descriptions of the waveforms based on the duty cycle and/or crest factor should be used to enable comparison and categorization of the qualitative waveform labels from different ESUs (