developed for 25(OH)D measurements, and was published in 1971 [41]. It used DBP as the binder and 3H-25(OH)D as tracer. Although the method was largely abandoned when efforts to streamline the assay by eliminating the extraction and purification procedures proved unsatisfactory, Roche Diagnostics has introduced an automated CPBA. The sample is incubated with ruthenium red labeled DBP to which 25(OH)D conjugated with biotin is then added to bind the free DBP. Streptavidin coated beads are added to bind the 25(OH)D biotin conjugate, the beads captured magnetically, and chemiluminescence induced. The concentration of 25(OH)D in the sample is inversely proportional to the chemiluminescence signal.

      Immunoassays. These fall into 3 main categories.

      Chemiluminescent assays were first commercialized by DiaSorin, although a number of such assays are now on the market. In these assays, the antibody is bound to a solid-phase material and sample 25(OH)D competes for binding with 25(OH)D conjugated to a chemiluminescent label. After washing the light signal is induced and quantitated. The amount of light produced is inversely related to the amount of 25(OH)D in the sample.

      1,25(OH)2D

      The measurement of 1,25(OH)2D is more challenging than that of 25(OH)D because it circulates in blood at levels that are nearly 1/1,000 that of 25(OH)D. As for the measurement of 25(OH)D, there are